Fluorescence Optical Detector Board

From Hackteria Wiki
Revision as of 07:32, 11 September 2020 by Gaudi (talk | contribs) (Sensor Spectrum)
Jump to: navigation, search

Learning from reverse engineering the Illumina Optical Detector Board extracted from a NeoPrep system (discontinued).

IMG 20200818 191434.jpg IMG 20200814 140900.jpg
The device electronics is quite straight forward. There is one line of 15 blue LEDs for excitation and two lines of photodiodes. The LEDs are controlled by an LED driver ICs and the photodiodes are read by an photo diode ADC. All comes together by a FPGA (black box). The power pins and input lines to the module are easy accessible on colour coded probe pins (nice).

Filters

The optical system for each of the two detector lines, is composed of 3 colour filters, 2 glass lens arrays, one plastic foil lens and 3 coated mirrors. IMG 20200814 133524.jpg IMG 20200814 140343.jpg


The 3 colour filters are:
Filter blue (excitation): 450 - 480 nm
Filter green (emission) : 510 - 550 nm
Filter red (emission) : 600 + nm


Mirrors

There are 3 mirrors in the system.

Mirror 1: transmission 440 - 510 nm (rest is reflected)
Mirror 2: seems to be a complete mirror
Mirror 3: is actually just a transparent glass

M145.png
Absorption spectrum from mirror 1.

It is not clear to me as to what the second line of photodiodes is measuring ass all the light seems to be reflected through mirror 2. (Maybe some infrared that I don't see or some very attenuated signal. Or maybe the second line is not used? confused).

Sensor Spectrum


The complete transmission spectrum of the first photo detector line with all the filters and mirrors in place:
540 nm + 5 / - 25 nm

File:TransmissionLine1.png
TransmissionLine1.png

LEDs

The excitation LED after filtering emits at 470 + 10 /- 15 nm

ExcitationLED withoutFilter.png
Excitation LED without filter.
ExcitationLED withFilter.png
Excitation LED with filter.

Luminex

Testing the sensor with "R&D Systems ® Luminex ® Bead-Based Assays for Multiplexing" kits.

IMG 20200909 222850.jpg
There are two light sources used to collect data from samples. The red light source (laser for Luminex® 100/200™ and LED for MAGPIX®) excites the internal dyes in the microspheres so that they can be classified on the bead map. The green light source excites any reporter bound to the surface of the beads (PE-Streptavidin).

In Luminex applications, SA-PE is excited by the 532 nm green laser or 520 nm green LED.
Its emission peak exists at 575 ± 10 nm (orange/yellow).

Systems with two LEDs:
(511 nm) green
(621 nm) red

PE-Streptavidin:
Excitation: 488 nm blue (azure) (broad excitation, one peak 488 nm, one at 575 nm)
Emission: 575 nm (yellow)

Beads:
Excitation 621 nm red ?
Emission? two dyes for fluorescence? Infrared?